User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Slaughter 103MP:HV Leakage Tester DATE: 29-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 23 NUMBER OF LINES: 287 CONFIGURATION: HP 34401A STANDARD: Yokogawa 2262 STANDARD: Clarostat 240-C ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD Yokogawa 2262 1.004 STD Clarostat 240-C 1.005 HEAD * INITIAL CONDITIONS * 1.006 DISP WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Set the UUT front panel controls as follows: 1.007 DISP ON/OFF switch ...................................... OFF 1.007 DISP AUTO/OFF/MAN switch ................................ OFF 1.007 DISP METER MULTIPLIER switch ........................... OFF 1.007 DISP REAL/TOTAL CURRENT switch ................ TOTAL CURRENT 1.008 DISP Mechanically zero both UUT panel meters. 1.009 DISP Set the UUT REJECTION INDICATORS (red limit pointers) 1.009 DISP on both panel meters to full scale or greater. 1.010 HEAD {* VOLTAGE ACCURACY TEST *} 1.011 JMP 2.001 1.012 EVAL 2.001 DISP [32] Connect the test equipment as follows: 2.001 DISP AC OUT HI (red) .................... 2262 6600V tap 2.001 DISP UUT GROUND (black) ................. 2262 V tap 2.001 DISP 2262 110V tap ...................... 34401 HI 2.001 DISP 2262 v tap ......................... 34401 LO 2.002 DISP Set the UUT front panel controls as follows: 2.002 DISP ON/OFF switch ....................................... ON 2.003 IEEE [@34401]FUNC "VOLT:AC" 2.004 IEEE [@34401]DET:BAND 20[GTL] 2.005 DISP To adjust the UUT output voltage, press and hold down 2.005 DISP the AUTO/OFF/MAN switch to the MAN position and 2.005 DISP momentarily position the INCREASE/DECREASE switch to 2.005 DISP INCREASE until the output reaches the required voltage. 2.005 DISP 2.005 DISP Momentarily position the INCREASE/DECREASE switch to 2.005 DISP DECREASE to lower the UUT output voltage. 2.006 DISP Adjust the TEST VOLTAGE control for a UUT reading 2.006 DISP of 2500 VAC. 2.007 ACC 2500.0V 0.24% # 2.007 IEEE [@34401]READ?[I] 2.008 IEEE [@34401]READ?[I][GTL] 2.009 MEM* 66 2.010 MEME 2.011 MEMC V 120.0U 60H #! Test Tol 120, Sys Tol 6, TUR 20.0000 (>= 4.00). 3.001 IEEE [@34401]FUNC "VOLT:AC" 3.002 IEEE [@34401]DET:BAND 20[GTL] 3.003 DISP Adjust the TEST VOLTAGE control for a UUT reading 3.003 DISP of 2000 VAC. 3.004 ACC 2000.0V 0.24% 3.005 IEEE [@34401]READ?[I] 3.006 MEM* 66 3.007 MEME 3.008 MEMC V 120.0U 60H #! Test Tol 120, Sys Tol 4.8, TUR 25.0000 (>= 4.00). 4.001 IEEE [@34401]FUNC "VOLT:AC" 4.002 IEEE [@34401]DET:BAND 20[GTL] 4.003 DISP Adjust the TEST VOLTAGE control for a UUT reading 4.003 DISP of 1500 VAC. 4.004 ACC 1500.0V 0.24% 4.005 IEEE [@34401]READ?[I] 4.006 MEM* 66 4.007 MEME 4.008 MEMC V 120.0U 60H #! Test Tol 120, Sys Tol 3.6, TUR 33.3333 (>= 4.00). 5.001 IEEE [@34401]FUNC "VOLT:AC" 5.002 IEEE [@34401]DET:BAND 20[GTL] 5.003 DISP Adjust the TEST VOLTAGE control for a UUT reading 5.003 DISP of 1000 VAC. 5.004 ACC 1000.0V 0.24% 5.005 IEEE [@34401]READ?[I] 5.006 MEM* 66 5.007 MEME 5.008 MEMC V 120.0U 60H #! Test Tol 120, Sys Tol 2.4, TUR 50.0000 (>= 4.00). 6.001 IEEE [@34401]FUNC "VOLT:AC" 6.002 IEEE [@34401]DET:BAND 20[GTL] 6.003 DISP Adjust the TEST VOLTAGE control for a UUT reading 6.003 DISP of 500 VAC. 6.004 ACC 500.0V 0.24% 6.005 IEEE [@34401]READ?[I] 6.006 MEM* 66 6.007 MEME 6.008 MEMC V 120.0U 60H #! Test Tol 120, Sys Tol 1.2, TUR 100.000 (>= 4.00). 7.001 DISP Release the AUTO/OFF/MAN switch to the OFF position. 7.002 DISP Set the UUT front panel controls as follows: 7.002 DISP ON/OFF switch ...................................... OFF 7.002 DISP METER MULTIPLIER switch ............................ X1 7.003 DISP Disconnect the UUT from the Yokogawa 2262. 7.004 HEAD {* TOTAL CURRENT MILLIAMMETER TEST *} 7.005 HEAD {* MULTIPLIER X1 *} 7.006 JMP 8.003 7.007 EVAL 8.001 DISP Disconnect the UUT from the Clarostat 240-C. 8.002 JMP 8.004 8.003 DISP Connect the HP 34401A HI/LO Input to the left hand 8.003 DISP RED/BLACK binding posts on the Clarostat 240-C. 8.004 DISP Connec the HP 34401A HI/LO 4W Sense/Ratio Ref 8.004 DISP inputs to the right hand 240-C binding post pair. 8.005 DISP Set the 240-c to 1,0,0,0,0,0. 8.006 34401 100kZ N 4W 8.007 MATH m[1] = mem * 1000 8.008 DISP Set the 240-c to 2,0,0,0,0,0. 8.009 34401 100kZ N 4W 8.010 MATH m[2] = mem * 1000 8.011 DISP Set the 240-c to 3,0,0,0,0,0. 8.012 34401 100kZ N 4W 8.013 MATH m[3] = mem * 1000 8.014 DISP Set the 240-c to 0,0,5,0,0,0. 8.015 34401 100kZ N 4W 8.016 MATH m[4] = mem * 1000 8.017 DISP Set the 240-c to 0,0,0,5,0,0. 8.018 34401 100kZ N 4W 8.019 MATH m[5] = mem * 1000 8.020 DISP Disconnect the 34401A Sense lines from the 240-C 8.020 DISP right side binding posts. 8.021 DISP Connect the UUT HIGH VOLTAGE output to the 240-C 8.021 DISP right side binding posts. 8.022 DISP Set the 240-C decade switches to 1, 0, 0, 0, 0, 0. 8.023 DISP Set the UUT front panel controls as follows: 8.023 DISP ON/OFF switch ....................................... ON 8.024 IEEE [@34401]FUNC "VOLT:AC" 8.025 IEEE [@34401]DET:BAND 20 8.026 DISP Adjust the TEST VOLTAGE control for a UUT reading 8.026 DISP of 5 mA AC. 8.027 IEEE [@34401]READ?[I] 8.028 MATH mem = 1000 * (mem / m[1]) 8.029 ACC 5.000mA 0.105% 8.030 MEME 8.031 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.00525, TUR 47.6190 (>= 4.00). 9.001 IEEE [@34401]FUNC "VOLT:AC" 9.002 IEEE [@34401]DET:BAND 20 9.003 DISP Adjust the TEST VOLTAGE control for a UUT reading 9.003 DISP of 4 mA AC. 9.004 IEEE [@34401]READ?[I] 9.005 MATH mem = 1000 * (mem / m[1]) 9.006 ACC 4.000mA 0.117% 9.007 MEME 9.008 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.00468, TUR 53.4188 (>= 4.00). 10.001 IEEE [@34401]FUNC "VOLT:AC" 10.002 IEEE [@34401]DET:BAND 20 10.003 DISP Adjust the TEST VOLTAGE control for a UUT reading 10.003 DISP of 3 mA AC. 10.004 IEEE [@34401]READ?[I] 10.005 MATH mem = 1000 * (mem / m[1]) 10.006 ACC 3.000mA 0.135% 10.007 MEME 10.008 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.00405, TUR 61.7284 (>= 4.00). 11.001 IEEE [@34401]FUNC "VOLT:AC" 11.002 IEEE [@34401]DET:BAND 20 11.003 DISP Set the 240-C decade switches to 2, 0, 0, 0, 0, 0. 11.004 DISP Adjust the TEST VOLTAGE control for a UUT reading 11.004 DISP of 2 mA AC. 11.005 IEEE [@34401]READ?[I] 11.006 MATH mem = 1000 * (mem / m[2]) 11.007 ACC 2.000mA 0.117% 11.008 MEME 11.009 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.00234, TUR 106.838 (>= 4.00). 12.001 IEEE [@34401]FUNC "VOLT:AC" 12.002 IEEE [@34401]DET:BAND 20 12.003 DISP Set the 240-C decade switches to 3, 0, 0, 0, 0, 0. 12.004 DISP Adjust the TEST VOLTAGE control for a UUT reading 12.004 DISP of 1 mA AC. 12.005 IEEE [@34401]READ?[I] 12.006 MATH mem = 1000 * (mem / m[3]) 12.007 ACC 1.000mA 0.135% 12.008 MEME 12.009 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.00135, TUR 185.185 (>= 4.00). 13.001 HEAD {* MULTIPLIER X10 *} 13.002 JMP 14.001 13.003 EVAL 14.001 DISP Release the AUTO/OFF/MAN switch to the OFF position. 14.002 DISP Set the 240-C decade switches to 0, 0, 5, 0, 0, 0. 14.003 DISP Set the UUT front panel controls as follows: 14.003 DISP METER MULTIPLIER switch ........................... X10 14.004 IEEE [@34401]FUNC "VOLT:AC" 14.005 IEEE [@34401]DET:BAND 20 14.006 DISP Adjust the TEST VOLTAGE control for a UUT reading 14.006 DISP of 50 mA AC. 14.007 IEEE [@34401]READ?[I] 14.008 MATH mem = 1000 * (mem / m[4]) 14.009 ACC 50.00mA 0.15% 14.010 MEME 14.011 MEMC mA 2.50U #! Test Tol 2.5, Sys Tol 0.075, TUR 33.3333 (>= 4.00). 15.001 DISP Release the AUTO/OFF/MAN switch to the OFF position. 15.002 HEAD {* MULTIPLIER X50 *} 15.003 JMP 16.001 15.004 EVAL 16.001 DISP Set the 240-C decade switches to 0, 0, 0, 5, 0, 0. 16.002 DISP Set the UUT front panel controls as follows: 16.002 DISP METER MULTIPLIER switch ........................... X50 16.003 IEEE [@34401]FUNC "VOLT:AC" 16.004 IEEE [@34401]DET:BAND 20 16.005 DISP Adjust the TEST VOLTAGE control for a UUT reading 16.005 DISP of 200 mA AC (4 on the milliammeter scale). 16.006 IEEE [@34401]READ?[I] 16.007 MATH mem = 1000 * (mem / m[5]) 16.008 ACC 200.00mA 0.091% 16.009 MEME 16.010 MEMC mA 12.50U #! Test Tol 12.5, Sys Tol 0.182, TUR 68.6813 (>= 4.00). 17.001 DISP Release the AUTO/OFF/MAN switch to the OFF position. 17.002 HEAD {*REAL CURRENT MILLIAMMETER TEST *} 17.003 HEAD {* MULTIPLIER X1 *} 17.004 JMP 18.001 17.005 EVAL 18.001 DISP Set the UUT front panel controls as follows: 18.001 DISP REAL/TOTAL CURRENT switch ................. REAL CURRENT 18.001 DISP METER MULTIPLIER switch ............................ X1 18.002 DISP Set the 240-C decade switches to 1, 0, 0, 0, 0, 0. 18.003 IEEE [@34401]FUNC "VOLT:AC" 18.004 IEEE [@34401]DET:BAND 20 18.005 DISP Adjust the TEST VOLTAGE control for a UUT reading 18.005 DISP of 5 mA AC. 18.006 IEEE [@34401]READ?[I] 18.007 MATH mem = 1000 * (mem / m[1]) 18.008 ACC 5.000mA 0.105% 18.009 MEME 18.010 MEMC mA 0.250U #! Test Tol 0.25, Sys Tol 0.00525, TUR 47.6190 (>= 4.00). 19.001 HEAD {* MULTIPLIER X10 *} 19.002 JMP 20.001 19.003 EVAL 20.001 DISP Release the AUTO/OFF/MAN switch to the OFF position. 20.002 DISP Set the 240-C decade switches to 0, 0, 5, 0, 0, 0. 20.003 DISP Set the UUT front panel controls as follows: 20.003 DISP METER MULTIPLIER switch ........................... X10 20.004 IEEE [@34401]FUNC "VOLT:AC" 20.005 IEEE [@34401]DET:BAND 20 20.006 DISP Adjust the TEST VOLTAGE control for a UUT reading 20.006 DISP of 50 mA AC. 20.007 IEEE [@34401]READ?[I] 20.008 MATH mem = 1000 * (mem / m[4]) 20.009 ACC 50.00mA 0.15% 20.010 MEME 20.011 MEMC mA 2.50U #! Test Tol 2.5, Sys Tol 0.075, TUR 33.3333 (>= 4.00). 21.001 DISP Release the AUTO/OFF/MAN switch to the OFF position. 21.002 HEAD {* MULTIPLIER X50 *} 21.003 JMP 22.001 21.004 EVAL 22.001 DISP Set the 240-C decade switches to 0, 0, 0, 5, 0, 0. 22.002 DISP Set the UUT front panel controls as follows: 22.002 DISP METER MULTIPLIER switch ........................... X50 22.003 IEEE [@34401]FUNC "VOLT:AC" 22.004 IEEE [@34401]DET:BAND 20 22.005 DISP Adjust the TEST VOLTAGE control for a UUT reading 22.005 DISP of 200 mA AC (4 on the milliammeter scale). 22.006 IEEE [@34401]READ?[I] 22.007 MATH mem = 1000 * (mem / m[5]) 22.008 ACC 200.00mA 0.091% 22.009 MEME 22.010 MEMC mA 12.50U #! Test Tol 12.5, Sys Tol 0.182, TUR 68.6813 (>= 4.00). 23.001 IEEE [@34401]FUNC "VOLT:AC" 23.002 IEEE [@34401]DET:BAND 20 23.003 DISP Release the AUTO/OFF/MAN switch to the OFF position. 23.004 DISP Set the UUT front panel controls as follows: 23.004 DISP ON/OFF switch ...................................... OFF 23.005 DISP Disconnect the test setup. 23.006 DISP end #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.